Manual control procedure – Audioscan Verifit User Manual

Page 59

Advertising
background image

Verifit

®

User's Guide Version 3.12

May 2015

3. Highlight and <PICK> [Battery type] and select from drop-down list.

4. Highlight and <PICK> [Hours per day] and select from drop-down list.

5. Highlight and <PICK> [Start test]

Manual control procedure

Manual control displays the reference microphone SPL, the coupler microphone SPL, their difference in dB
(gain) and coupler microphone signal distortion at any selected 1/12th octave frequency and at a selected input
level from 40 to 90 dB (in 5 dB steps) and at 0 dB (off).

1. Press <Tests>, then highlight and <PICK> [Test box measures], then highlight and <PICK> [Manual control]

2. Highlight and <PICK> the Level window, then highlight and <PICK> the desired stimulus level from the

drop-down list.

3. Highlight and <PICK> the Frequency window, then highlight and <PICK> the desired frequency from the

drop-down list. (Distortion measurement data is available only from 200-4000 Hz.)

4. Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the

measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the
test.