Verifit

®

User's Guide Version 3.12

May 2015

2. Press <Tests>, then highlight and <PICK> [On-ear measures], then highlight and <PICK> [Manual control].

3. Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB stimulus level from the drop-

down list.

4. Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12th octave, A-weighted, or C-

weighted from the drop-down list.

5. For 1/12th octave filter, highlight and <PICK> the Frequency window, then highlight and <PICK> the

desired frequency from the drop-down list.

6. Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the

measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the
test.