On-ear manual control – Audioscan Verifit User Manual

Page 75

Advertising
background image

Verifit

®

User's Guide Version 3.12

May 2015

Refer to On-Ear Measures - Setup to prepare the system for use and for proper positioning of the client and the
probe tube.

1. Press <Tests>, then highlight and <PICK> [On-ear measures], then highlight and pick [Noise reduction].

2. Highlight and <PICK> Test 1, 2 or 3.

3. Highlight and <PICK> a stimulus type from the list box.

4. Highlight and <PICK> a stimulus level from the list box.

5. Highlight and <PICK> [Start test]

6. Highlight and <PICK> [Continue] to capture the comparison curve once the noise reduction function has

responded.

On-ear manual control

1. Refer to On-Ear Measures - Setup to prepare the system for use and for proper positioning of the client and

the probe tube.

2. Press <Tests>, then highlight and <PICK> [On-ear measures], then highlight and <PICK> [Manual control].

75

Advertising