Audioscan Verifit User Manual

Page 83

Advertising
background image

Verifit

®

User's Guide Version 3.12

May 2015

thresholds measured with insert phones.

For threshold corrections in the HL-to-SPL transform, the Verifit uses an HA-1 RECD or an HA-2 RECD
including tubing effects, based on whether the stated HL transducer is Insert+foam or Insert+mold
respectively. If the RECD was measured with an earmold, these tubing effects will be specific to the client’s
earmold; if it is measured with a foam tip, they will be based on average earmold tubing.

For Test box Speechmap involving BTE instruments the earmold tubing must be considered. If the BTE is
coupled to the HA-1 coupler using the personal earmold, the tubing effects will be part of the measurement.
If the HA-2 (BTE) coupler is used, the HA-1 referenced RECD will be converted to an HA-2 RECD. As
with the threshold correction above, the use of the earmold in RECD will include the effects specific to the
client’s earmold, otherwise an average earmold is assumed.

The following describes acceptable RECD measurement protocols and Test-box Speechmap coupling
combinations for BTE instruments and the implications of each in the test box simulation.

ANSI S3.46 method:

RECD On-ear measurement: Audioscan RECD transducer + foam eartip

Test-box Speechmap coupling: client’s personal earmold + HA-1 (ITE) coupler

With this method, RECD is a property of the ear only, and not dependent on the transducer or sound source
coupling. The earmold effects will be part of the measurement in the test box. Measured RECD is appropriate
for coupler simulation and HL threshold correction, and error is minimized.

Alternate method #1:

RECD On-ear measurement: Audioscan RECD transducer + client’s personal earmold

Test-box Speechmap coupling: HA-2 (BTE) coupler

This method utilizes the client’s personal earmold for the on-ear RECD measurement. The HA-1 RECD is
converted to HA-2 reference with earmold effects retained and is applied in Testbox Speechmap.The error
introduced is the difference between the earmold effects when connected to the hearing instrument receiver
and earmold effects when connected to the RECD transducer.

If the total earmold tubing length is less than 35mm, it is recommended that Foam tip be selected in the
RECD Coupling field. This will bypass the correction and increase accuracy for small earmolds.

Alternate method #2:

RECD On-ear measurement: Audioscan RECD transducer + foam eartip

Test-box Speechmap coupling: HA-2 (BTE) coupler

Since the RECD here is measured with a foam eartip, an average earmold is added to obtain an HA-2
RECD with earmold effects. The error involved will be the difference between the actual earmold and the
average used.

Alternate method #3:

RECD On-ear measurement: Audioscan RECD transducer + client’s personal earmold

83

Advertising
This manual is related to the following products: