Reur measurement procedure – Audioscan Verifit User Manual

Page 87

Advertising
background image

Verifit

®

User's Guide Version 3.12

May 2015

barcode scanner and Input device operation.

To enter data from a barcode see Barcode data input and:

1. Press the barcode trigger and center the red line on the barcode from a distance of 6 – 8 in. (15 – 20 cm). The

scanner will beep and the Barcode entry poster opens.

2. Highlight and <PICK> [Done] to accept the data indicated on the poster.

To enter assessment data and parameters manually:

1. Highlight and <PICK> [Audiometry].

2. Highlight and <PICK> [Transducer], then highlight and <PICK> from the list.

3. Highlight and <PICK> [REUR], then highlight and <PICK> to toggle selection.

Measured REUR should be used with a soundfield threshold because individual REUR is part of the threshold.
Average REUR should be used with headphone or insert phone thresholds because individual REUR isn’t part of
these thresholds.

4. When all selections have been completed, press <Continue>.

5. Enter HL threshold values using a mouse or keypad. Left click the mouse on the audiogram form to enter a

point. Click again to delete the point. On the keypad use the arrow keys to change frequency and level and
press <PICK> to enter or delete a threshold point.

6. Press <Continue> when all points have been entered.

The HL for the highest frequency is used for aided targets at higher frequencies.

REUR measurement procedure

Performing a REUR test erases all existing REAR and REIG data associated with the current test ear. REUR
should be measured only if Transducer = Soundfield. Otherwise it is more accurate to use average. See
Audiometric data entry.

1. Press <Tests>, then highlight and <PICK> [On-ear measures], then highlight and <PICK> [Insertion gain].

2. Referring to On-Ear Measures - Setup and On-Ear Measures Screen Setup, prepare the system, position the

patient and insert the probe tube.

3. If REUR is set to Average, highlight and <PICK> [Audiometry] then highlight and <PICK> [REUR] to

toggle to [Measure]. Press <Continue> to advance to the Threshold entry poster. You may enter threshold
data at this time (see Audiometric data entry) or press <Continue> to proceed with the measurement.

87