Seaward Supernova XE User Manual
Page 27
How to perform Automatic Tests
17
Default Sequence
When you first enter this menu all of the previous values are
displayed and one of the field descriptions is highlighted.
Use the cursor keys to highlight a field to be changed and press
Enter to select the desired option.
Test Parameter Fields
By selecting one of the test parameter fields from the Default
---------------DEFAULT SEQUENCE EDITOR----------------
OUTPUT LIMIT DUR NUM
SOCKET
230V
VISUAL
YES
1
EARTH BOND
4A 0.10
Ω 5 1
INSULATION
500V 0.50M
Ω 5 1
SUB LEAKAGE
SKIP
LEAKAGE
SKIP
TOUCH LEAKAGE
SKIP
POLARITY
SKIP
PERFORM TEST
-------------------------------LEAKAGE--------------------------------
OUTPUT
LIMIT
DURATION
NUMBER
230V
SKIP
2s
1
115V
0.25mA
5s
2
0.50mA
10s 3
0.75mA
30s
4
1.50mA
60s
5
2.25mA
120s 10
2.50mA
180s
15
3.00mA
UL
20
3.50mA
25
9.90mA
30