Barcodes, Interfacing – Seaward Supernova XE User Manual

Page 74

Advertising
background image

Barcodes

64

Barcodes

Barcode Scanner Specification

The Supernova XE Tester can be used with barcodes and barcode
scanners, readers or wands compatible with the Seaward PAT2000i.
These connect to the serial (RS232) connector The required
configuration is as follows:-

Baud Rate:

9600

Start Bits:

2

Data Bits

8

Stop Bits:

2

Parity:

None

Inter-character Delay: 20ms

When to use a Barcode Scanner

Barcodes can be scanned at the following points:-

• Entering a Testcode as part of an Automatic Test

- Test Setting Editor

- Use Testcodes

• Entering an Appliance number

Interfacing

The Tester provides two interfacing ports:-

Serial Port - for connection to a PC, printer or Barcode Reader

To avoid any problems during download, ensure that leads are
undamaged and correctly wired.

The Connector pinouts for Serial port are shown below:-