Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-54

Table 2-28 Average Number of T-Power Measurement

Parameter

Description

Default

10

Range

1 to 1000

Unit

N/A

Knob Step

1

Direction Key Step

1


2. Avg Mode

Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
 When “Exp” is selected, the result is the exponential average of the

current N (N is specified in Avg Num) measurement results.

 When “Repeat” is selected, the result is the arithmetic average of the

current N (N is specified in Avg Num) measurement results.

3. TP Type

Peak

Display the power of the signal with the maximum amplitude between the
start line and stop line. The detector type is set to “Pos Peak”
automatically.

Average

Display the average power of signals between the start line and stop line.
The detector type is set to “Voltage Avg” automatically.

RMS

Display the mean square root (in power unit) of the voltages of signals
between the start line and stop line. The detector type is set to “RMS
Avg” automatically.

4. Start Line

Set the left margin (in time unit) of T-Power measurement. The data
calculated under this measurement is between the start line and stop line.
You can use the numeric keys, knob or direction keys to modify this
parameter.