Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-66

Table 2-41 Average Number for EBW Measurement

Parameter

Description

Default

10

Range

1 to 1000

Unit

N/A

Knob Step

1

Direction Key Step

1


2. Avg Mode

Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
 When “Exp” is selected, the result is the exponential average of the

current N (N is specified in Avg Num) measurement results.

 When “Repeat” is selected, the result is the arithmetic average of the

current N (N is specified in Avg Num) measurement results.


3. Max Hold

Enable or disable max hold and the default is “Off”.
 When Max Hold is enabled, each measurement result is compared with

the previous result and the maximum is displayed.

 When Max Hold is disabled, the current measurement result is displayed.
 Max Hold and average measurement mode are mutually exclusive and

average measurement mode will be automatically disabled when Max
Hold is enabled.

4. EBW Span

This span which is the same with the span of the analyzer is the frequency
range for the sweep. Modifying this parameter will also change the span of
the analyzer. You can use the numeric keys, knob or direction keys to modify
this parameter.

Table 2-42 Span for EBW Measurement

Parameter

Description

Default

2 MHz

Range

100 Hz to 7.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

EBW Span/100, the minimum is 1 Hz

Direction Key Step

in 1-1.5-2-3-5-7.5 sequence