Chapter 5 Specifications

RIGOL

User’s Guide for DSA800 Series

5-5

Displayed Average Noise Level (DANL)

DSA832

DSA875

Frequency

attenuation = 0 dB, RBW = VBW = 10 Hz, sample
detector, trace average ≥ 50, tracking generator off,
20℃ to 30℃, input impendence = 50 Ω

PA
off

9 kHz to 100 kHz <-110 dBm (typ.)

<-110 dBm (typ.)

100 kHz to 5 MHz

<-125 dBm,
<-128 dBm (typ.)

<-125 dBm,
<-128 dBm (typ.)

5 MHz to 3.2 GHz

<-130 dBm,
<-134 dBm (typ.)

<-130 dBm,
<-134 dBm (typ.)

3.2 GHz to 6 GHz

<-126 dBm,
<-130 dBm (typ.)

6 GHz to 7.5 GHz

<-121 dBm,
<-125 dBm (typ.)

PA
on

100 kHz to 1 MHz <-142 dBm (typ.)

<-142 dBm (typ.)

1 MHz to 5 MHz

<-142 dBm,
<-145 dBm (typ.)

<-142 dBm,
<-145 dBm (typ.)

5 MHz to 3.2 GHz

<-147 dBm,
<-151 dBm (typ.)

<-147 dBm,
<-151 dBm (typ.)

3.2 GHz to 6 GHz

<-143 dBm,
<-147 dBm (typ.)

6 GHz to 7.5 GHz

<-138 dBm,
<-142 dBm (typ.)