Marker, Mkr->cf, Mkr->step – RIGOL DSA875 User Manual

Page 137: Mkr->start, Marker-> -85

Advertising
background image

Chapter 2 Front Panel Operation

RIGOL

User’s Guide for DSA800 Series

2-85

Marker->


Set the other system parameters (such as center frequency and reference level)
using the current marker readings. Press Marker-> to enable a marker
automatically if no marker is currently enabled.

Mkr->CF


Set the center frequency of the analyzer to the frequency of the current marker.
If Normal marker is selected, the center frequency will be set to the

frequency of the current marker.

If Delta, Delta Pair, or Span Pair marker is selected, the center frequency

will be set to the frequency of the Delta Marker.

The function is invalid in Zero span mode.

Mkr->Step


Set the center frequency step of the analyzer to the frequency of the current
marker.
If Normal marker is selected, the center frequency step will be set to the

frequency of the current marker.

If Delta, Delta Pair or Span Pair marker is selected, the center frequency

step will be set to the frequency of the Delta Marker.

The function is invalid in Zero span mode.

Mkr->Start


Set the start frequency of the analyzer to the frequency of the current marker.
If Normal marker is selected, the start frequency will be set to the frequency

of the current marker.

If Delta, Delta Pair or Span Pair marker is selected, the start frequency will

be set to the frequency of the Delta Marker.

The function is invalid in Zero span mode.