Cf step, Signal track – RIGOL DSA875 User Manual

Page 58

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-6

CF Step


Set the step of center frequency. Changing the center frequency in a fixed step
continuously switches the channel to be measured.

Key Points:
 The CF step can be set in “Manual” or “Auto” mode. In Auto mode, the CF step

is 1/10 of the span in Non-zero span mode or equals the RBW while in Zero
span mode; in Manual mode, you can set the step using the numeric keys.

 After you set an appropriate CF step and select Center Freq, use the up and

down direction keys to switch between measurement channels in a specified
step in order to sweep the adjacent channel manually.

 You can modify this parameter using the numeric keys, knob or direction keys.

For more details, please refer to “Parameter Setting”.

Table 2-5 CF Step

Parameter

Explanation

Default

750 MHz

Range

1 Hz to 7.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

Span > 0, step = Span/200
Span = 0, step = 100 Hz
Min = 1 Hz

Direction Key Step in 1, 2, 5 sequence


Signal Track


Turn on or off signal track. This function is used to track and measure signal with
unstable frequency and less than 3 dB transient variation in amplitude. Place
Marker1 (see Marker Measurements) onto the signal under measurement to
track and measure the variation of the signal continuously.

The signal track process is as shown in the figure below: