RIGOL DSA875 User Manual

Page 108

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-56

ACP


Measurement Interface:

Figure 2-15 Adjacent Channel Power Measurement Interface


Measurement Results: Main CH Pwr, Upper and Lower.
 CH Pwr: display the power within the bandwidth of the main channel.
 Upper: display the power of the upper channel and the power difference

between the upper channel and the main channel (in dBc).

 Lower: display the power of the lower channel and the power difference

between the lower channel and the main channel (in dBc).


Measurement Parameters: average number, average mode, main channel
bandwidth, adjacent channel bandwidth and channel spacing.

1. Avg Num

Specify the number of averages used when calculating the measurement
result and the default is “Off”. You can use the numeric keys, knob or direction
keys to modify this parameter.



Main Channel Bandwidth

Adjacent Channel Bandwidth

Channel Spacing