Sweep and function settings, Bw/det, Sweep and function settings -23 – RIGOL DSA875 User Manual

Page 75: Bw/det -23

Advertising
background image

Chapter 2 Front Panel Operation

RIGOL

User’s Guide for DSA800 Series

2-23

Sweep and Function Settings

BW/Det


Set the RBW (Resolution Bandwidth), VBW (Video Bandwidth) and detector type
parameters of the analyzer.

RBW


Set the desired resolution bandwidth in order to distinguish between signals which
are close in frequency.

Key Points:
 Reducing RBW will increase the frequency resolution, but will increase the

sweep time (Sweep Time is affected by a combination of RBW and VBW when
it is in Auto mode).

 RBW varies with the span (non-zero span) in Auto RBW mode.
 You can modify this parameter using the numeric keys, knob or direction keys.

For more details, please refer to Parameter Setting.


Note: If the detector type is “Quasi-Peak” or the filter type is “EMI”, RBW can be
200 Hz, 9 kHz or 120 kHz only.

Table 2-13 RBW (Filter Type is Gauss)

Parameter

Explanation

Default

1 MHz

Range

10 Hz to 1 MHz

Unit

GHz, MHz, kHz, Hz

Knob Step

in 1, 3, 10 sequence

Direction Key Step in 1, 3, 10 sequence


Advertising