Amplitude

Measurement Range

Range

f

c

≥ 10 MHz

DANL to +20 dBm

Maximum Input Level
DC voltage

50 V

CW RF power

attenuation = 30 dB
+20 dBm (100 mW)

Max. damage level*

+30 dBm (1 W)

Displayed Average Noise Level (DANL)

DSA815

Frequency

attenuation = 0 dB, RBW = VBW = 100 Hz, sample
detector, trace average ≥ 50, tracking generator off,
20℃ to 30℃, input impendence = 50 Ω

PA
off

100 kHz to 1 MHz <-90 dBm, <-110 dBm (typ.)
1 MHz to 1.5 GHz <-110 dBm + 6 × (f/1 GHz) dB, <-115 dBm (typ.)

PA
on

100 kHz to 1 MHz <-110 dBm, <-130 dBm (typ.)
1 MHz to 1.5 GHz <-130 dBm + 6 × (f/1 GHz) dB, <-135 dBm (typ.)















Note: *When f

c

≥ 10 MHz, input level > +25 dBm and PA is Off, the protection switch will be on.