Sweep/trig, Sweep time, Sweep/trig -28 – RIGOL DSA875 User Manual

Page 80

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-28

Sweep/Trig


Set parameters about the Sweep and Trigger functions, including sweep time, auto
sweep time, sweep mode, number of sweep, trigger type, etc.

Sweep Time


Set the time needed for the spectrum analyzer to finish a sweep within the span
range. The sweep time can be set in “Auto” or “Manual” mode and the default is
“Auto”.

Key Points:
 In non-zero span, the analyzer selects the shortest sweep time on the basis of

the current RBW and VBW settings if Auto is selected.

 Decreasing the sweep time would speed the measurement. However, an error

may be caused if the specified sweep time is less than the minimum sweep
time in Auto coupling; at this point, “UNCAL” is shown in the status bar on the
screen.

 You can modify this parameter using the numeric keys, knob or direction keys.

For more details, please refer to Parameter Setting.


Table 2-16 Sweep Time

Parameter

Explanation

Default

37.5 ms

Range*

20 us to 7500 s

Unit

ks, s, ms, us, ns, ps

Knob Step

Sweep time/100, Min = 1 us

Direction Key Step

in 1, 1.5, 2, 3, 5, 7.5 sequence

Note: *The minimum in non-zero span mode is 1 ms.