Troubleshooting, Troubleshooting -2 – RIGOL DSA875 User Manual

Page 194

Advertising
background image

RIGOL

Chapter 4 Troubleshooting&Message

User’s Guide for DSA800 Series

4-2

Troubleshooting


The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still, please contact

RIGOL and provide your device

information (SystemInformationSystem Info).

1
The screen is still dark (no display) after power on:

(1) Check whether the fan is running:

If yes, the connection of the internal screen cable may be loose.

If not, the instrument has failed to start up and please refer to step (2).

(2) Check the power:

Check whether the power supply has been connected correctly and the

power switch has been turned on.

Check whether the power fuse is burned. If a new fuse needs to be

installed, please use a specified fuse ( 250V AC, T2A; 5 mm×20 mm).


2. The key is unresponsive or gives a wrong response:

(1) Press all the keys at the front panel to check if all of them are normal after

power on.

(2) Press System

Self-Test

Key Test to check if all the keys are

working properly.

(3) If a key is not working, the numeric keyboard connection might be loose or

the numeric keyboard is broken. Do not disassemble the instrument by
yourself and contact

RIGOL.


3. The spectrum lines on the screen do not update for a long period of

time:
(1) Check whether the screen is locked; if so, press Esc to unlock it.
(2) Verify whether all the trigger conditions have been met and whether there

is a valid trigger signal.

(3) Check whether the analyzer is in single sweep.
(4) Check whether the current sweep time is too long.



Advertising