Chapter 4 Troubleshooting&Message

RIGOL

User’s Guide for DSA800 Series

4-9

Error Message

Command Error

Number Message Description
-100

Command error.
A generic syntax error indicates that the device cannot detect more
specific errors. This code indicates that only Command Error defined in
IEEE 488.2,11.5.1.1.4 has been found.

-101

Invalid character.
The syntactic element contains an invalid character; for example, the
command header contains an ampersand (SETUP&). This error might
be used in place of errors -114, -121, -141 and perhaps some others.

-102

Syntax error.
An unrecognizable command or data type was encountered; for
example, a string not supported by the device was received.

-103

Invalid separator.
The parser was expecting a separator but recieved an illegal character;
for example, the semicolon following a program message was omitted:
*EMC 1:CH1:VOLTS 5.

-104

Data type error.
The parser recognized a data element other than the one allowed; for
example, numeric or string data was expected but block data was
encountered.

-105

GET not allowed.
A Group Execute Trigger was contained in a program message (see
IEEE 488.2, 7.7).

-108

Parameter not allowed.
More parameters were received than expected for the command
header; for example, the *EMC common command only accepts one
parameter, so *EMC 0, 1 is not allowed.

-109

Missing parameter.
Fewer parameters were contained than required in the command
header; for example, the *EMC common command requires one
parameter, so *EMC is not allowed.

-110

Command header error.

Advertising