Chapter 2 Front Panel Operation

RIGOL

User’s Guide for DSA800 Series

2-69

Table 2-44 Average Number for C/N Ratio Measurement

Parameter

Description

Default

10

Range

1 to 1000

Unit

N/A

Knob Step

1

Direction Key Step

1


2. Avg Mode

Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
 When “Exp” is selected, the result is the exponential average of the

current N (N is specified in Avg Num) measurement results.

 When “Repeat” is selected, the result is the arithmetic average of the

current N (N is specified in Avg Num) measurement results.


3. Offset Freq

Set the difference between the carrier center frequency and the noise center
frequency. You can use the numeric keys, knob and direction keys to modify
this parameter.

Table 2-45 Frequency Offset for C/N Ratio Measurement

Parameter

Description

Default

2 MHz

Range

33 Hz to 2.5 GHz

Unit

GHz, MHz, kHz, Hz

Knob Step

Offset Frequency/100, the
minimum is 1Hz

Direction Key Step

in 1-1.5-2-3-5-7.5 sequence

4. Noise BW

Set the bandwidth of the noise to be measured. You can use the numeric keys,
knob or direction keys to modify this parameter.