Chapter 4 Troubleshooting&Message

RIGOL

User’s Guide for DSA800 Series

4-19

359

Invalid function when trace is not enabled.
A marker could not be assigned to a trace when the trace is not enabled.

360

User preset failed.
The user preset has failed because the edition of the saved status data is
incorrect or the status data has been damaged. In this situation, the
system will execute Factory settings as the defaults.

400

Input signal power out of range.
Indicate that the input signal power is out of range. Working in this state
for a long time would cause damage to the device. Please reduce the
input power.

401

1st LO unlock.
The first LO is unlocked. Please report the error to your

RIGOL sales or

technical support.

402

2nd LO unlock.
The second LO is unlocked. Please report the error to your

RIGOL sales

or technical support.

403

Track LO unlock.
Track LO is unlocked. Please report the error to your

RIGOL sales or

technical support.

412

DA overrange due to volume adjustment in FM.

413

Intermediate frequency signal out of range.

420

Option not installed.
The desired operation cannot be performed because the option required
is not installed.

460

No peak found.
No signal peak that meets the criteria was found under Peak Search.

461

No fundamental wave found.
Indicate that no fundamental wave greater than 50dBm is found.

462

No two-tone signal found.

463

Invalid function as cursor is not enabled.
The cursor cannot be set to the reference level unless it is enabled.

464

Menus under Storage can not be set to Userkey.

465

Invalid function.

470

Calibration memory lost.
Indicate that the calibration data before delivery or last valid calibration
data has been lost. Please report the error to your

RIGOL sales or

technical support team.