Input atten, Scale/div – RIGOL DSA875 User Manual

Page 66

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-14

Table 2-7 Reference Level

Parameter

Explanation

Default

0 dBm

Range

-100 dBm to 20 dBm

Unit

dBm, -dBm, mV, uV

Knob Step

In Log scale mode, step = Scale/10
In Lin scale mode, step = 0.1 dBm

Direction Key Step In Log scale mode, step = Scale

In Lin scale mode, step = 1 dBm


Input Atten


Set the front attenuator of the RF input in order to ensure big signals (or small
signals) to pass through the mixer with low distortion (or low noise).

Key Points:
 When the preamplifier is turned on, the input attenuation could be set up to

30 dB. You can adjust the reference level to ensure that the specified
parameter meets the inequality (2-3).

 You can modify this parameter using the numeric keys, knob or direction keys.

For more details, please refer to Parameter Setting.

Table 2-8 Input Attenuation

Parameter

Explanation

Default

10 dB

Range

0 dB to 30 dB

Unit

dB

Knob Step

1 dB

Direction Key Step 5 dB


Scale/Div


Set the logarithmic units per vertical grid division on the display. This function is
only available when the scale type is set to “Log”.

Advertising
This manual is related to the following products: