Single – RIGOL DSA875 User Manual

Page 82

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-30

Enter continuous sweep mode and

wait for trigger condition

Sweep again and wait for trigger

condition

Select

Cont

Single sweep status?

Y

N

Figure 2-9 Process of Continuous Sweep


Single


In single sweep mode, this menu is used to execute trigger initialization. After that,
the analyzer will execute the specified number of sweeps (or measurements) if the
trigger conditions are satisfied.

Key Points:
 If the instrument is in continuous sweep mode and no measurement function

is enabled, press this key and the system will enter single sweep mode and
execute the specified number of sweeps if the trigger conditions are satisfied.

 If the instrument is in continuous sweep mode and a measurement function is

on, press this key and the system will enter single measurement mode and
execute the specified number of measurements if the trigger conditions are
satisfied.

 If the system is already in single sweep mode, press this key and the system

will execute the specified number of sweeps (or measurements) if the trigger
conditions are satisfied.

 In single sweep mode, trigger initialization (press Sweep/TrigSingle or

send the “:INIT” command through the remote interface) should be executed
before trigger condition judgment.