Tg lvl offset, Power sweep (only for dsa815) – RIGOL DSA875 User Manual

Page 94

Advertising
background image

RIGOL

Chapter 2 Front Panel Operation

User’s Guide for DSA800 Series

2-42

TG Lvl Offset


Assign a certain offset to the output power of the TG when gains or losses occur
between the TG output and external device in order to display the actual power
value.

Key Points:
 This parameter only changes the readout of the TG output power, rather than

the actual value of it.

 The offset could be either a positive (gain in the external output) or a negative

(loss in the external output).

 You can use the numeric keys, knob or direction keys to modify this parameter.

For more details, please refer to Parameter Setting.


Table 2-23 TG Level Offset

Parameter

Explanation

Default

0 dB

Range

-200 dB to 200 dB

Unit

dB

Knob Step

1 dB

Direction Key Step 10 dB


Power Sweep (Only for DSA815)


Enable or disable the power sweep function.

When enabled, the output power of the TG varies with the sweep rate of the
analyzer within the specified sweep frequency range (from start to stop frequency)
and increases gradually from the specified TG power (see TG Level) within the
specified power range (see Power Range). When disabled, the TG output is
fixed at the specified power (see TG Level).

Note: This function in not available in zero span mode.