Mode 1 input timing, Mode 1 input timing -6 – National Instruments Data Acquisition Device E Series User Manual

Page 101

Advertising
background image

Chapter 4

Digital I/O

E Series User Manual

4-6

ni.com

Mode 1 Input Timing

(NI 6016 and NI 6025E Devices Only)

Figure 4-3 and Table 4-3 show timing

specifications for an input transfer in mode 1.

Figure 4-3. Input Transfer in Mode 1 Timing Specifications

Table 4-3. Input Transfer in Mode 1 Timing Specifications

Name

Description

Minimum (ns)

Maximum (ns)

T1

STB* Pulse Width

100

T2

STB* = 0 to IBF = 1

150

T3

Data before STB* = 1

20

T4

STB* = 1 to INTR = 1

150

T5

Data after STB* = 1

50

T6

RD* = 0 to INTR = 0

200

T7

STB* Pulse Width

150

T1

T2

T4

T7

T6

T3

T5

STB*

IBF

INTR

RD*

DATA

Advertising