Figure a-22. daqpad-6020e block diagram, Daqpad-6020e block diagram, Differential signals – National Instruments Data Acquisition Device E Series User Manual

Page 161

Advertising
background image

Appendix A

Device-Specific Information

E Series User Manual

A-26

ni.com

DAQPad-6020E Block Diagram

Figure A-22 shows a block diagram of the DAQPad-6020E.

Figure A-22. DAQPad-6020E Block Diagram

Connecting Signals to the DAQPad-6020E BNC

Analog Input

You can use each analog input BNC connector for one differential signal or
two single-ended signals.

Differential Signals

To connect differential signals, determine the type of signal source you are
using: a floating signal source or a ground-referenced signal source. Refer
to the

Differential Connection Considerations

and

Connecting Analog

Input Signals

sections of Chapter 2,

Analog Input

, for more information.

To measure a floating signal source, move the switch to the FS position. To
measure a ground-referenced signal source, move the switch to the GS

Timing

PFI / Trigger

I/O Connector

4

USB Connector

Exter

nal P

o

w

e

r

Digital I/O (8)

12-Bit

Sampling

A/D

Converter

EEPROM

Configuration

Memory

+
NI-PGIA
Gain
Amplifier

Calibration

Mux

Mux Mode
Selection
Switches

Analog

Muxes

Voltage

REF

Calibration

DACs

Dither

Circuitry

6

Calibration

DACs

5 V

Supply

DAC0

DAC1

1 A Fuse

DAQ - STC

Analog Input

Timing/Control

Analog Output

Timing/Control

Digital I/O

Trigger

Counter/

Timing I/O

Interrupt
Request

Bus

Interface

(8)

(8)

1

2

AI Control

Analog

Input

Control

EEPROM

Control

DAQ-PnP

DAQ-STC

Bus

Interface

Analog

Output

Control

Bus

Interface

Bus

Interface

USB

Port

USB Micro

Controller

Interrupt
Request

EEPROM

Control

IRQ

AO Control

Data (16)

Data (16)

Data

Transceivers

ADC

FIFO

Advertising