Appendix C

Troubleshooting

E Series User Manual

C-2

ni.com

reference the signal to the same ground level as the device reference. There
are various methods of achieving this reference while maintaining a high
common-mode Rejection Ratio (CMRR). These methods are outlined in
the

Connecting Analog Input Signals

section of Chapter 2,

Analog Input

.

AI GND is an AI common signal that routes directly to the ground
connection point on the devices. You can use this signal if you need a
general analog ground connection point to the device. Refer to the

Differential Connection Considerations

section of Chapter 2,

Analog

Input

, for more information.

How can I use the AI Sample Clock and AI Convert Clock signals on
an E Series device to sample the AI channel(s)?

E Series devices use the ai/SampleClock and ai/ConvertClock signals to
perform interval sampling. As Figure C-1 shows, ai/SampleClock controls
the sample period, which is determined by the following equation:

1/sample period = sample rate

Figure C-1. Interval Sample

The ai/ConvertClock signal controls the convert period, which is
determined by the following equation:

1/convert period = convert rate

This method allows multiple channels to be sampled relatively quickly in
relationship to the overall scan rate, providing a nearly simultaneous effect
with a fixed delay between channels.

Channel 0

Channel 1

Convert Period

Sample Period