Figure 2-21. reference trigger final buffer, Using a digital source, Using a digital source -35 – National Instruments Data Acquisition Device E Series User Manual

Page 68

Advertising
background image

Chapter 2

Analog Input

© National Instruments Corporation

2-35

E Series User Manual

When the reference trigger occurs, the DAQ device continues to write
samples to the buffer until the buffer contains the desired number of
posttrigger samples. Figure 2-21 shows the final buffer.

Figure 2-21. Reference Trigger Final Buffer

Using a Digital Source

To use ai/ReferenceTrigger with a digital source, specify a source and an
edge. The source can be an external signal connected to any PFI or
RTSI <0..6> pin. The source can also be one of several internal signals on
your DAQ device. Refer to Device Routing in MAX in the NI-DAQmx Help
or the LabVIEW 8.x Help for more information.

Also, specify whether the measurement acquisition stops on the rising edge
or falling edge of the ai/ReferenceTrigger signal.

Figure 2-22 shows the timing requirements of the ai/ReferenceTrigger
source.

Figure 2-22. ai/ReferenceTrigger Source Timing Requirements

Pre-Trigger Samples

Reference Trigger

Post-Trigger Samples

Complete Buffer

Rising-Edge

Polarity

Falling-Edge

Polarity

t

w

t

w

= 10 ns minimum