position. Figure A-23 shows the source type switch locations on the front
panel of the BNC DAQPads.

Figure A-23. BNC DAQPad Front Panel

Figure A-24 shows the analog input circuitry on BNC DAQPads. When the
switch is in the FS position, AI x – is grounded through a 0.1

μF capacitor

in parallel with a 5 k

Ω resistor.

Figure A-24. BNC DAQPads Analog Input Circuitry

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

1

3

5

7

9 11 13 15 17 19 21 23 25 27 29

2

4

6

8 10 12 14 16 18 20 22 24 26 28 30

USER 2

USER 1

PWR

COM

CTR 0

OUT

EXT
REF

DAC 0

OUT

ACH 6

ACH 4

ACH 2

ACH 0

ACH 1

ACH 3

ACH 5

ACH 7

DAC 1

OUT

PFI 0/

TRIG 1

FS

GS

FLOATING SOURCE

GND REF SOURCE

FS

GS

Source Type

Switch

Source Type

Switch

FS = Floating Source
GS = Grounded Source

AI

x +

AI

x –

GS

0.1 µF

5 k

Ω

AI GND

FS