Table 2-6. recommended input configurations – National Instruments Data Acquisition Device E Series User Manual

Page 54

Advertising
background image

Chapter 2

Analog Input

© National Instruments Corporation

2-21

E Series User Manual

Refer to the

Analog Input Terminal Configuration

section for descriptions

of the input modes.

Table 2-6. Recommended Input Configurations

Input

Signal Source Type

Floating Signal Sources

(Not Connect To Building

Ground)

Ground-Referenced Signal

Sources

Examples

• Ungrounded thermocouples

• Signal conditioning with

isolated outputs

• Battery devices

Examples

• Plug-in instruments

with non-isolated outputs

Differential (DIFF)

Single-Ended—
Ground Referenced
(RSE)

Single-Ended—
Non-Referenced
(NRSE)

+

+

Vs

AI –

AI GND

R

ext

AI +

+

+

V

s

AI +

AI –

AI GND

+

+

V

s

AI

AI GND

+

+

V

1

AI

+ V

g

AI GND

Ground-loop losses, Vg, are added

to measured signal.

NOT RECOMMENDED

+

+

V

s

AI

AI SENSE

AI GND

AI GND

R

ext

+

+

Vs

AI

AI SENSE

AI GND