National Instruments Data Acquisition Device E Series User Manual

Page 77

Advertising
background image

Chapter 2

Analog Input

E Series User Manual

2-44

ni.com

Figure 2-32. ai/SampleClock and ai/ConvertClock Signals

ai/ConvertClock

ai/SampleClock

1

2

3

Sample Clock too fast for Convert Clock.
Sample Clock pulses are gated off.

ai/ConvertClock

ai/SampleClock

1

2

3

Convert Clock too fast for Sample Clock.
Convert Clock pulses are gated off.

ai/ConvertClock

ai/SampleClock

1

2

3

Improperly matched Sample Clock and Convert Clock.
Leads to aperiodic sampling.

ai/ConvertClock

ai/SampleClock

1

2

3

Properly matched Sample Clock and Convert Clock.

Advertising