Appendix A

Device-Specific Information

© National Instruments Corporation

A-73

E Series User Manual

Single-Ended Signals

For each BNC connector that you use for two single-ended channels, set the
source type switch to the GS position. This setting disconnects the built-in
ground reference resistor from the negative terminal of the BNC connector,
allowing the connector to be used as a single-ended channel, as shown in
Figure A-59.

Figure A-59. Single-Ended Signals

When you set the source type to the GS position and software-configure the
device for single-ended input, each BNC connector provides access to
two single-ended channels, AI x and AI x+8. For example, the BNC
connector labeled AI 0 provides access to single-ended channels AI 0 and
AI 8, the BNC connector labeled AI 1 provides access to single-ended
channels AI 1 and AI 9, and so on. Up to 16 single-ended channels are
available in single-ended measurement modes.

For a detailed description of each signal, refer to the

I/O Connector Signal

Descriptions

section of Chapter 1,

DAQ System Overview

.

Analog Output

You can access analog output signals on the BNC connectors labeled AO 0
and AO 1. Figure A-60 shows the analog output circuitry on BNC
DAQPads.

Figure A-60. Analog Output Circuitry

AI

x + 8

AI

x

GS

0.1 µF

5 k

Ω

AI GND

FS

AO

AO GND