AI Hold Complete Event Signal

AI Hold Complete Event (ai/HoldCompleteEvent) is an output-only signal
that generates a pulse with the leading edge occurring approximately 50 to
100 ns after an A/D conversion begins. The polarity of this output is
software-selectable, but is typically configured so that a low-to-high
leading edge can clock external AI multiplexers indicating when the input
signal has been sampled and can be removed. This signal has a 400 to
500 ns pulse width and is software-enabled. Figure 2-34 shows the timing
for ai/HoldCompleteEvent.

Figure 2-34. ai/HoldCompleteEvent Timing

External Strobe Signal

External Strobe is an output-only signal on the EXT STROBE pin that
generates either a single pulse or a sequence of eight pulses in the
hardware-strobe mode. An external device can use this signal to latch
signals or to trigger events. In the single-pulse mode, software controls the
level of External Strobe. A 10 ms and a 1.2

μs clock are available for

generating a sequence of eight pulses in the hardware-strobe mode.
Figure 2-35 shows the timing for the hardware-strobe mode External
Strobe signal.

Figure 2-35. External Strobe Timing

Note

External Strobe is used for signal conditioning with SCXI and is not available for

use with NI-DAQmx.

t

w

= 400 to 500 ns

t

d

= 50 to 100 ns

ai/ConvertClock

ai/HoldCompleteEvent

t

d

t

w

t

w

= 600 ns or 5

μs

t

w

t

w

V

OH

V

OL

Advertising