Appendix A

Device-Specific Information

E Series User Manual

A-2

ni.com

NI 6011E (NI PCI-MIO-16XE-50) Dither

You cannot disable dither on the NI 6011E (NI PCI-MIO-16XE-50). The
ADC resolution on this device is so fine that the ADC and the PGIA
inherently produce almost 0.5 LSB

rms

of noise. This configuration is

equivalent to having a dither circuit that is always enabled.

NI 6011E (NI PCI-MIO-16XE-50) Block Diagram

Figure A-1 shows a block diagram of the NI 6011E
(NI PCI-MIO-16XE-50).

Figure A-1. NI 6011E (NI PCI-MIO-16XE-50) Block Diagram

PCI Bus

Configuration

Memory

Timing

PFI / Trigger

I/O Connector

2

Digital I/O (8)

16-Bit

Sampling

A/D

Converter

Programmable
Gain
Amplifier

Calibration

Mux

Mux Mode

Selection

Switches

Voltage

REF

Calibration

DACs

4

Calibration

DACs

DAC0

DAC1

DAQ - STC

Analog Input

Timing/Control

Analog Output

Timing/Control

Digital I/O

Trigger

Counter/

Timing I/O

RTSI Bus

Interface

DMA/

Interrupt
Request

Bus

Interface

(8)

(8)

AI Control

IRQ

DMA

AO Control

Data (16)

EEPROM

Address/Data

Control

Data (16)

Analog

Input

Control

EEPROM

Control

DMA

Interface

MIO

Interface

DAQ-STC

Bus

Interface

Analog

Output

Control

I/O

Bus

Interface

Address (5)

Analog

Muxes

ADC

FIFO

MITE

Generic

Bus

Interface

PCI
Bus

Interface

+

Data (16)

RTSI Bus

2