Chapter 4 Using Power Software
Configuration tool
Setting up Communications
PowerSmart Socket Power Quality Meter
57
3. From the “Look in” box, select the directory where a
new database will be stored. By default, it will be the
“Sites” directory. Type a site name for your device in
the “File name” box, click New, and then click OK.
4. On the Instrument Setup tab, select “PSS” in the
“Model” box. Power Software automatically selects the
appropriate instrument options for your meter.
5. Select a correct CT secondary current (5A or 1A) for
your meter.
6. If you wish to add any comments for your meter, type
them into the “Comment” box.
Setting up Communications
You can communicate with your meters via a PC RS-232
serial port, through the Internet, via either a local Ethernet,
or a wireless GPRS Dial-Up connection, and via the USB port.
To
configure
your
communications
with
the
PowerSmart Socket PQM:
1. Select Configuration from the Tools menu. Under the
Communication group on the Instrument Setup tab,
select the type of connection for your device.
2. Set the device communication address you assigned to
the
PowerSmart
Socket
PQM
port.
When
communicating via the Ethernet or a GPRS modem,
the PowerSmart Socket PQM responds to any address
you select.
3. In the “Sampling Rate” box, select a rate at which
Power Software updates data on your screen when you
continuously poll the device in the Power Software
Data Monitor.
The communication protocol and port settings must match
the settings you made in your meter.