Appendix III: SMA Protocol
A. STANDARD SCALE RESPONSE MESSAGE
Most of the host commands are responded to in the following message format. The only
host commands that do not are the:
D
iagnostic, ABout and INformation commands
<LF> <s> <r> <n> <m> <f> <xxxxxx.xxx> <uuu> <CR>
where:
<LF>
Start of response message
<s>
scale status
definition / example
‘Z’
Center of Zero <xxxxxx.xxx>= 0.000
‘O’
Over
Capacity
<xxxxxx.xxx>=
+weight
‘U’
Under
Capacity <xxxxxx.xxx>= -weight
‘E’
Zero Error (clears when condition clears)
‘I’
Initial-Zero Error (if used, this error is
maintained until zero condition is cleared)
‘T’
Tare Error (clears after being read)
<space>
None of the above conditions
Note: For ‘E’, ‘I’, ‘T’ error conditions
<xxxxxx.xxx>= ————— (center dashes)
and ‘Z’, ‘O’, ‘U’ are overridden.
<r>
range
(‘1’, ‘2’, ‘3’, etc.) always ‘1’ for single range
<n>
gross/net status
‘G’
Gross normal weight
‘T’
Tare weight (in response to ‘M’ command)
‘N’
Net normal weight
‘g’
gross weight in high-resolution
‘n’
net weight in high-resolution
<m>
motion status
‘M’
scale in Motion
<space>
scale not in Motion
<f>
future
reserved for future or custom use
<xxxxxx.xxx>
weight data this field is fixed at 10 characters
<uuu>
Unit of Measure
<CR>
End of response message