<2. Handling Cautions>

2-10

IM 01S01C01-01EN

2.8.3 CENELEC ATEX Certification

(1) Technical Data

a. CENELEC ATEX (DEKRA) Intrinsically Safe

Type

Caution for CENELEC ATEX (DEKRA)

Intrinsically safe type.

Note 1. FVX110 Fieldbus Segment Indicator

with optional code /KS25 for potentially

explosive atmospheres:

• No. DEKRA 11ATEX0022 X

• Applicable Standard: EN 60079-0:2009,

EN 60079-11:2007, EN 60079-26:2007,

EN 60079-27:2008, EN 61241-11:2006

Note 2. Ratings

Type of Protection and Marking Code:

II1G Ex ia IIB/IIC T4 Ga

II1D Ex ia IIIC T80ºC Da IP6X

Group: II

Category: 1G 1D

Ambient Temperature: –40* to 60ºC

* –15ºC when O-ring material is Fluoro-rubber.

Maximum Surface Temperature for dust-proof.

T80ºC (Tamb.: –40* to 60ºC)

* –15ºC when O-ring material is Fluoro-rubber.

Degree of Protection of the Enclosure:

IP66 and IP67

Electrical Data

• When combined with Trapezoidal output

characteristic FISCO model IIC or IIB barrier

[Supply/Output circuit (terminals + and –)]

Ui = 17.5 V, Ii = 500 mA, Pi = 5.5 W,

Ci = 3.52 nF, Li = 0 µH

• When combined with Linear characteristic

barrier

[Supply/Output circuit (terminals + and –)]

Ui = 24.0 V, Ii = 250 mA, Pi = 1.2 W,

Ci = 3.52 nF, Li = 0 µH

Note 3. Installation

• All wiring shall comply with local installation

requirements. (Refer to the installation

diagram)

Note 4. Maintenance and Repair

• The instrument modification or parts

replacement by other than authorized

representative of Yokogawa Electric

Corporation is prohibited and will void

DEKRA Intrinsically safe Certification.

Note 5. Special Conditions for Safe Use

• In the case where the enclosure of the

segment indicator is made of aluminium, if

it is mounted in an area where the use of

category 1 G apparatus is required, it must

be installed such, that even in the event of

rare incidents, ignition sources due to impact

and friction sparks are excluded.

Note 6. Installation instructions

• The test voltage for the isolation between the

intrincically safe supply/output circuit and the

frame of the apparatas for segment indicator

that are provided with surge protection is

limited to 90 V, due to the presence of the

surge protection device only.

When used in a potentially explosive

atmosphere, requiring the use of apparatus

of equipment category 1D or 2D, certified

cable entry devices shall be used that are

suitable for the application and correctly

installed.

● FISCO Model

Non-Hazardous

Locations

Hazardous Locations

F0209.ai

Terminator

(FISCO Model)

Ex i

Field Instruments

(Passive)

Hand-

held-

Terminal

Supply Unit and

Safety Barrier

(FISCO Model)

Terminator

Data

U

U

I

I.S. fieldbus system complying with FISCO