Yokogawa FVX110 User Manual

Page 76

Advertising
background image

<12. Parameter Lists>

12-3

IM 01S01C01-01EN

Relative

Index Index

Parameter Name

Factory Default Write

Mode

Explanation

54

1054

SOFTDWN_FORMAT 0x01

AUTO Selects the software download method.

0x01: Standard

0x02: YOKOGAWA Standard

55

1055

SOFTDWN_COUNT

0

Indicates the number of times the internal FlashROM was

erased.

56

1056

SOFTDWN_ACT_

AREA

0

Indicates the ROM number of the currently working FlashROM.

0: FlashROM #0 working

1: FlashROM #1 working

57

1057

SOFTDWN_MOD_

REV

1, 0, 0, 0, 0, 0,

0, 0, 0

Indicates the software module revision.

58

1058

SOFTDWN_ERROR 0

Indicates the error during a software download.

Refer Table A8.4.

59

1059 SOFTDWN_HISTORY

Not used by the FVX110.

60

1060 SOFTDWN_HIST_INDEX

0

AUTO

Not used by the FVX110.

61

1061 COMPATIBILITY_REV 1

Indicates the smallest Rev value compatible with device

DevRev.

62

1062 CAPABILITY_LEV

0x00

Indicates the capability level of instrument interior.

63

1063 CAPABILITY_CONFIG

0x0000

A parameter corresponding to AP_CONF or DEV_OPTIONS

before the addition of parameter CAPABILITY_LEV.

64

1064 WRITE_LOCK_LEVEL

2

AUTO

Specifies blocks that activates Write Lock.

65

1065 SI_CONTROL_CODES

1

A parameter for switching to make the instrument compatible

with SI units.

66

1066 FD_VER

0

Indicates value of major version of instrument diagnostics

specifications (FF-912).

67

1067 FD_FAIL_ACTIVE

0

A parameter that corresponds to ”Failed” in the NAMUR NE-

107 category.

68

1068 FD_OFFSPEC_ACTIVE 0

A parameter that corresponds to ”Off Specification” in the

NAMUR NE-107 category.

69

1069 FD_MAINT_ACTIVE

0

A parameter that corresponds to ”Maintenance” in the NAMUR

NE-107 category.

70

1070 FD_CHECK_ACTIVE 0

A parameter that corresponds to ”Check Function” in the

NAMUR NE-107 category.

71

1071 FD_FAIL_MAP

0xFC000000

AUTO

Specifies the bit assigned to FD_FAIL_ACTIVE, a parameter

for indicating ”Failed,” a 32-bit alarm listed in FD_SIMULATE.

DiagnosticValue.

72

1072 FD_OFFSPEC_MAP

0x00003800

AUTO

Specifies the bit assigned to FD_OFFSPEC_ACTIVE, a

parameter for indicating ”Off Specification,” a 32-bit alarm listed

in FD_SIMULATE. DiagnosticValue.

73

1073 FD_MAINT_MAP

0x000003E0

AUTO

Specifies the bit assigned to FD_MAINT_ACTIVE, a parameter

for indicating ”Maintenance,” a 32-bit alarm listed in FD_

SIMULATE. DiagnosticValue.

74

1074 FD_CHECK_MAP

0x01FF8008

AUTO

Specifies the bit assigned to FD_CHECK_ACTIVE, a

parameter for indicating ”Check Function,” a 32-bit alarm listed

in FD_SIMULATE. DiagnosticValue.

75

1075 FD_FAIL_MASK

0xFFFFFFFF

AUTO

Specifies the bit that notifies the host of 32-bit ”Failed” alarms

listed in FD_FAIL_ACTIVE.

76

1076 FD_OFFSPEC_MASK 0xFFFFFFFF

AUTO

Specifies the bit that notifies the host of 32-bit ”Off

Specification” alarms listed in FD_OFFSPEC_ACTIVE.

77

1077 FD_MAINT_MASK

0xFFFFFFFF

AUTO

A parameter that specifies the bit that notifies the host of

32-bit ”Maintenance” alarms listed in FD_MAINT_ACTIVE. A

parameter set by the user.

78

1078 FD_CHECK_MASK

0xFFFFFFFF

AUTO

Specifies the bit that notifies the host of 32-bit ”Check Function”

alarms listed in FD_CHECK_ACTIVE.

79

1079 FD_FAIL_ALM

AUTO

Indicates alarm information for alarms categorized under

”Failed.”

80

1080 FD_OFFSPEC_ALM

AUTO

Indicates alarm information for alarms categorized under ”Off

Specification.”

81

1081 FD_MAINT_ALM

AUTO

Indicates alarm information for alarms categorized under

”Maintenance”.

82

1082 FD_CHECK_ALM

AUTO

Indicates alarm information for alarms categorized under

”Check Function”.

83

1083 FD_FAIL_PRI

0

AUTO

Indicates the FD_FAIL_ALM priority for an alarm.

84

1084 FD_OFFSPEC_PRI

0

AUTO

Indicates the FD_OFFSPEC_ALM priority for an alarm.

85

1085 FD_MAINT_PRI

0

AUTO

Indicates theFD_MAINT_ALM priority for an alarm.

Advertising