Index – Yokogawa PC-Based MX100 User Manual

Page 101

Advertising
background image

4-13

IM MX100-01E

Specification

3

2

1

4

5

Index

The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).

Input

Measurement

Range Type

Rated Measurement

Range

Measurement Accuracy

Highest

Resolution

(1 Digit)

Integration Time: 16.67 ms

or More

Integration Time: 1.67 ms

DC voltage

60 mV (high res.)

0.000 to 60.000 mV

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

1 µV

1 V

–1.0000 to 1.0000 V

±(0.05% of rdg + 2 digits)

±(0.1% of rdg + 10 digits)

100 µV

6 V (high res.)

0.0000 to 6.0000 V

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

100 µV

Thermocouple

(excludes RJC accuracy

, when

burnout is OFF)

PLATINEL

0.0 to 1400.0°C

±(0.05% of rdg + 1°C)

±(0.1% of rdg + 4°C)

0.1°C

PR40-20

0.0 to 1900.0°C

±(0.05% of rdg + 2.5°C)

Except 300 to 700°C: ±6°C

less than 300°C: not guaranteed

±(0.1% of rdg + 12°C)

Except 300 to 700°C: ±25°C

less than 300°C: not guaranteed

NiNiMo

0.0 to 1310.0°C

±(0.05% of rdg + 0.7°C)

±(0.1% of rdg + 2.7°C)

WRe3-25

0.0 to 2400.0°C

±(0.05% of rdg + 2°C)

Except 0 to 200°C: ±2.5°C

2000°C or higher:

±(0.05% of rdg + 4°C)

±(0.1% of rdg + 7°C)

Except 0 to 200°C: ±12°C

2000°C or higher:

±(0.1% of rdg + 11°C)

W/WRe26

0.0 to 2400.0°C

±(0.05% of rdg + 2°C)

Except 100 to 300°C: ±4°C

less than 100°C: not guaranteed

±(0.1% of rdg + 8.5°C)

Except 100 to 300°C: ±12°C

less than 100°C: not guaranteed

N (AWG14)

0.0 to 1300.0°C

±(0.05% of rdg + 0.7°C)

±(0.1% of rdg + 3.5°C)

XK GOST

–200.0 to 600.0°C

±(0.05% of rdg + 0.5°C)

Except –200 to 0°C:

±(0.2% of rdg + 0.7°C)

±(0.1% of rdg + 2.5°C)

Except –200 to 0°C:

±(1% of rdg + 2.5°C)

RTD

(Measurement

current: 1 mA)

Cu10 at 20°C

alpha = 0.00392

–200.0 to 300.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

0.1°C

Cu10 at 20°C

alpha = 0.00393

–200.0 to 300.0°C

Cu25 at 0°C

alpha = 0.00425

–200.0 to 300.0°C

±(0.1% of rdg + 0.5°C)

±(0.2% of rdg + 2°C)

Cu53 at 0°C

alpha = 0.00426035 –50.0 to 150.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

Cu100 at 0°C

alpha = 0.00425

–50.0 to 150.0°C

Pt25 (JPt100 × 1/4) –200.0 to 550.0°C

±(0.1% of rdg + 0.5°C)

±(0.2% of rdg + 2°C)

Cu10 GE (high res.) –200.0 to 300.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

Cu10 L&N (high res.) –200.0 to 300.0°C

Cu10 WEED

(high res.)

–200.0 to 300.0°C

Cu10 BAILEY

(high res.)

–200.0 to 300.0°C

Pt100 GOST

–200.0 to 600.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

Cu100 GOST

–200.0 to 200.0°C

Cu50 GOST

–200.0 to 200.0°C

Cu10 GOST

–200.0 to 200.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

Measurement interval, integration time, and filter:

Measurement

Interval

Integration

Time

Burnout

Detection Cycle

Filter

Rejected Noise and Notes

100 ms

1.67 ms

1 s

*1

Rectangular

600 Hz and its integer multiples

*2

200 ms

Measurement

interval

500 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply

frequency and set 16.67 or 20 ms

1 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

2 s

100 ms

Rectangular

10 Hz and its integer multiples

5, 10, 20, 30, 60 s 200 ms

Cos

Fc = 5-Hz low-pass filter

*1 When the measurement interval is 100 ms, burnout is detected in one channel per measurement interval.

Therefore, if measurement is started in a burnout condition or after a burnout occurs, burnout cannot be

detected for up to 10 measurements (approximately 1 second).

*2 Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially with

temperature measurement. In such cases, set the measurement interval to 500 ms or higher, or use the

4-CH, High-Speed Universal Input Module.

Reference junction compensation:

Switch between external and internal on each channel, remote
RJC function available

Reference junction compensation accuracy:

When measuring temperature greater than or equal to 0°C and
when the temperature of the input terminal is balanced

Type R, S, W: ±1°C

Type K, J, E, T, N, L, U, XK GOST: ±0.5°C

Type N(AWG14), PLATINEL, NiNiMo, WRe3-25, W/WRe26: ±1°C
Note: The internal reference junction compensation is fixed to 0°C

for type B and PR40-20

4.5 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10) Specifications

Advertising