2 main module (mx100-e) specifications, Main module (mx100-e) specifications -2, Measurement – Yokogawa PC-Based MX100 User Manual

Page 90: Computation

Advertising
background image

4-2

IM MX100-01E

4.2 Main Module (MX100-E) Specifications

Measurement

Style number:

S3

Meas. range and accuracy:

See the measurement range and accuracy given in the
specifications of each input module.

Max. number of inputs: 60 (however a maximum of six modules can be controlled)
Meas. interval:

Select from 10, 50, 100, 200, 500 ms, 1, 2, 5, 10, 20, 30, or 60 s.

Up to three intervals defined for the input modules can be set
(multi-interval)

Synch. between modules:

Synchronized within the same sampling interval (in the same unit)

Synch. between channels:

4-CH, High-Speed Universal Module and Digital Input Module:
synchronized between channels. 10-CH, Medium-Speed Universal
Input Module, 30-CH, Medium-Speed DCV/TC/DI Input Module,
6-CH, Medium-Speed Four-Wire RTD Resistance Input Module, and
the 4-CH, Medium-Speed Strain Input Module: due to the scanner
type, not synchronized between channels

Time setting:

Configured based on the time information sent from the PC at the
start of measurement

Filter function:

First-order lag filter, applicable measurement types:

DC voltage, thermocouple, RTD, strain, and resistance. You can select
a time constant (time until 63.2% of the output value is reached)
corresponding to the sampling interval indicated in the table below.

Time constant = measurement interval × N (where N = 5, 10, 20,
25, 40, 50, or 100)

Measurement

Interval (s)

Selectable Time Constants (s)

N = 5 N = 10 N = 20 N = 25 N = 40 N = 50 N = 100

0.01

0.05

0.1

0.2

0.25

0.4

0.5

1

0.05

0.25

0.5

1

1.25

2

2.5

5

0.1

0.5

1

2

2.5

4

5

10

0.2

1

2

4

5

8

10

20

0.5

2.5

5

10

12.5

20

25

50

1

5

10

20

25

40

50

100

2

10

20

40

50

80

100

200

5

25

50

100

125

200

250

500

10

50

100

200

250

400

500

1000

20

100

200

400

500

800

1000

2000

30

150

300

600

750

1200

1500

3000

60

300

600

1200

1500

2400

3000

6000

Computation

Computation performed on the main module
Difference comp. between channels:

Difference computation between arbitrary channels (DC voltage,
thermocouple, RTD, DI, strain, resistance, and scaling)

Linear scaling:

Possible scaling ranges: DC voltage, thermocouple, RTD, DI, strain,
and resistance

Scaling range: –30000 to 30000

Scaling display range: –32000 to 32000

Decimal point position: Arbitrary