Index – Yokogawa PC-Based MX100 User Manual

Page 109

Advertising
background image

4-21

IM MX100-01E

Specification

3

2

1

4

5

Index

The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).

Input

Measurement

Range Type

Rated Measurement

Range

Measurement Accuracy

Highest

Resolution

(1 Digit)

Integration Time: 16.67 ms

or More

Integration Time: 1.67 ms

DC voltage

60 mV (high res.)

0.000 to 60.000 mV

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

1 µV

1 V

–1.0000 to 1.0000 V

±(0.05% of rdg + 2 digits)

±(0.1% of rdg + 10 digits)

100 µV

6 V (high res.)

0.0000 to 6.0000 V

±(0.05% of rdg + 20 digits)

±(0.1% of rdg + 100 digits)

100 µV

4-wire RTD

(Measurement

current: 1 mA)

Cu10 at 20°C

alpha = 0.00392

–200.0 to 300.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

0.1°C

Cu10 at 20°C

alpha = 0.00393

–200.0 to 300.0°C

Cu25 at 0°C

alpha = 0.00425

–200.0 to 300.0°C

±(0.1% of rdg + 0.5°C)

±(0.2% of rdg + 2°C)

Cu53 at 0°C

alpha = 0.00426035 –50.0 to 150.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

Cu100 at 0°C

alpha = 0.00425

–50.0 to 150.0°C

Pt25 (JPt100 × 1/4) –200.0 to 550.0°C

±(0.1% of rdg + 0.5°C)

±(0.2% of rdg + 2°C)

Cu10 GE

(high resolution)

–200.0 to 300.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

Cu10 L&N

(high resolution)

–200.0 to 300.0°C

Cu10 WEED

(high resolution)

–200.0 to 300.0°C

Cu10 BAILEY

(high resolution)

–200.0 to 300.0°C

Pt100 GOST

–200.0 to 600.0°C

±(0.05% of rdg + 0.3°C)

±(0.1% of rdg + 1.5°C)

Cu100 GOST

–200.0 to 200.0°C

Cu50 GOST

–200.0 to 200.0°C

Cu10 GOST

–200.0 to 200.0°C

±(0.1% of rdg + 2°C)

±(0.2% of rdg + 5°C)

Measurement interval, integration time, and filter:

Measurement Interval Integration Time

Filter

Rejected Noise and Notes

100, 200 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

500 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply

frequency and set 16.67 or 20 ms

1 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

2 s

100 ms

Rectangular

10 Hz and its integer multiples

5,10, 20, 30, 60 s

200 ms

Cos

Fc = 5-Hz low-pass filter

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially

with temperature measurement and 20 Ω measurement. In such cases, set the measurement interval

to 500 ms or higher.

Maximum input voltage:

DC voltage at 1-V range or less, RTD, resistance, and DI (contact):
±10 VDC (continuous)

Other measurement ranges: ±120 VDC (continuous)

Normal-mode voltage:

DC voltage, DI (LEVEL): 1.2 times the range rating or less (50/60
Hz, peak value including the signal component). The RTD and
resistance ranges indicate the voltage conversion value when
measurement current applied.)

Resistance 2 kΩ, RTD 100 Ω, 500 Ω, 1000 Ω systems: 50 mV peak

Resistance 200 Ω, RTD 10 Ω, 25 Ω, 50 Ω systems: 10 mV peak

Resistance 20 Ω: 4 mV peak

Normal-mode rejection ratio:

RTD and registance ranges are voltage conversion values when
the measured current is flowing.

For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)

For integration time of 1.67 ms: 50/60 Hz is not rejected

4.7 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06) Specifications

Advertising