Chapter 3 troubleshooting and maintenance, Errors at power on, System errors – Yokogawa PC-Based MX100 User Manual

Page 73: Errors at power on -1 system errors -1, Index

Advertising
background image

3-1

IM MX100-01E

Troubleshooting and Maintenance

3

2

1

4

5

Index

Chapter 3

Troubleshooting and Maintenance

3.1 Error Display on the 7-Segment LED and

Corrective Actions

The main module has a two-digit 7-segment LED as shown in the figure below. The
7-segment LED displays the system status. This section describes the displays on the
7-segment LED when errors occur on the system and their corrective actions. For normal
displays other than errors, see “Contents Displayed on the 7-Segment LED” in section 1.2.

7-segment LED

DATA ACQUISITION UNIT

ETHERNET

POWER

10BASE - T

100BASE - TX

100 - 240V AC

Do not operate without reading safety precaution in users manual.

70VA MAX

50/60Hz

SW

ON

1 2 3 4 5 6 7 8

If servicing is necessary, or if the instrument is not operating correctly after performing
the corrective actions below, contact your nearest YOKOGAWA dealer.

Errors at Power ON

The left and right digits of the 7-segment LED display “b” and an error code, respectively.
The LED illuminates.

Probable Cause

Corrective Action

Ref. Page

The display is b* (where * is any character

other than F). The dip switch settings are not

correct.

Turn OFF the power, eject the CF card, turn ON all dip switches,

and start up again. If the situation does not change servicing is

required.

2-27

The display is bF.

The dip switch settings are not correct.

Powering up in setup reset mode. Turn OFF the power, turn ON

all dip switches, and power up again. Since all settings such as

the IP address are initialized, reconfiguration is necessary.

2-23

System Errors

The left and right digits of the 7-segment LED display “F” and an error code, respectively.
The LED illuminates.

Display Probable Cause

Corrective Action

Ref. Page

F0

System ROM error.

Servicing required.

-

F1

SRAM error

Servicing required.

-

F2

EEPROM error

Servicing required.

-

F3

Error in the internal battery of the

main module.

Servicing required.

However, this error is also displayed immediately after the battery

is replaced. If this happens, power-cycle the MX100.

-

F4

Ethernet controller error

Servicing required.

-

F5

Internal data error

(only when the /DS option is enabled)

Servicing required.

-

Advertising