On the MX100, the integral time is determined by the measurement interval setting as
shown in the table below.

4-CH, High-Speed Universal Input Module

Measurement

Interval

Integration

Time

Filter

Rejected Noise and Notes

10 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

50 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply frequency

and set 16.67 or 20 ms

100, 200 ms

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

500 ms

100 ms

Rectangular

10 Hz and its integer multiples

1, 2, 5,10, 20,

30, 60 s

200 ms

Cos

Fc = 5-Hz low-pass filter

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate

especially with temperature measurement. In such cases, set the measurement interval to 50

ms or higher.

10-CH, Middle-Speed Universal Input Module/6-CH, Medium-Speed Four-Wire RTD

Resistance Input Module

Measurement

Interval

Integration

Time

Filter

Rejected Noise and Notes

100, 200 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

500 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply frequency

and set 16.67 or 20 ms

1 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

2 s

100 ms

Rectangular

10 Hz and its integer multiples

5, 10, 20, 30, 60 s 200 ms

Cos

Fc = 5-Hz low-pass filter

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate

especially with temperature measurement. In such cases, set the measurement interval to 500

ms or higher, or use the 4-CH, High-Speed Universal Input Module

30-CH, Middle-Speed DCV/TC/DI Input Module

Measurement

Interval

Integration

Time

Filter

Rejected Noise and Notes

500 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

1 s

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply frequency

and set 16.67 or 20 ms

2 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

5, 10, 20, 30, 60 s 100 ms

Rectangular

10 Hz and its integer multiples

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate

especially with temperature measurement. In such cases, set the measurement interval to 1 s

or higher, or use the 4-CH, High-Speed Universal Input Module or the 10-CH, Medium-Speed

Universal Input Module.

4-CH, Medium-Speed Strain Input Module

Measurement

Interval

Integration

Time

Filter

Rejected Noise and Notes

100 ms

1.67 ms

Rectangular

600 Hz and its integer multiples*

200 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply frequency

and set 16.67 or 20 ms

500 ms

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

1 s

100 ms

Rectangular

10 Hz and its integer multiples

2, 5,10, 20,

30, 60 s

200 ms

Cos

Fc = 5-Hz low-pass filter

*

Since the power supply frequency noise is not rejected, the measured vales may fluctuate. In

such cases, set the measurement interval to 200 ms or higher.

2.7 Measures against Noise on the MX100