Output type, Output waveform, Pulse interval – Yokogawa PC-Based MX100 User Manual

Page 38: Output update interval, Output range, Operation when errors occur upon startup

Advertising
background image

1-29

IM MX100-01E

Explanation of Functions

3

2

1

4

5

Index

1.10 Functions of the 8-CH, Medium-Speed PWM

Output Module

This module is equipped with eight outputs that can output pulse wave duty. A duty pulse
waveform is output according to the specified pulse interval. A pulse interval can be set
for each channel.

Input terminal (clamp terminal)

Terminal cover

Output Type

Transmission output: Outputs a duty pulse waveform corresponding to the measurement

computation results of the input channel specified on the same unit.

User output:

The specified value is output based on the data sent from the PC
software. Transmission output of computation channel data or of
data from separate units is possible.

Output Waveform

Pulse interval

Output width

External

power supply

voltage

Duty=

Output width

Pulse interval

×100 [%]

Pulse Interval

1 ms to 300 s (can be set channel by channel)
However,
1 ms interval setting rng: 1 ms to 30.000 s

Can be set in units of 1 ms

10 ms interval setting rng: 10 ms to 300.00 s

Can be set in units of 10 ms

Output Update Interval

The output is updated every 100 ms (minimum). It is not synchronized to the
measurement interval.

Output Range

0.000% to 100.000%

Operation When Errors Occur upon Startup

See section 1.11.

Advertising