Yokogawa PC-Based MX100 User Manual

Page 98

Advertising
background image

4-10

IM MX100-01E

Reference junction compensation accuracy:

When measuring temperature greater than or equal to 0°C and
when the temperature of the input terminal is balanced

Type R, S, W: ±1°C

Type K, J, E, T, N, L, U, XK GOST: ±0.5°C

Type N(AWG14), PLATINEL, NiNiMo, WRe3-25, W/WRe26: ±1°C
Note: The internal reference junction compensation is fixed to 0°C

for type B and PR40-20

Maximum input voltage:

DC voltage at 1-V range or less, thermocouple, RTD, and
DI (contact): ±10 VDC (continuous)

Other measurement ranges: ±120 VDC (continuous)

Normal-mode voltage: DC voltage, thermocouple, DI (LEVEL): 1.2 times the range rating

or less (50/60 Hz, peak value including the signal component)

RTD 100 Ω: 50 mV peak

RTD 10 Ω, 25 Ω, 50 Ω: 10 mV peak

Normal-mode rejection ratio:

For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz ±0.1%)

For integration time of 1.67 ms: 50/60 Hz is not rejected

The RTD and resistance ranges indicates the voltage conversion
value when current flows.

Common-mode voltage: 600 VAC rms (50/60 Hz), reinforced (double) insulation
Common-mode rejection ratio:

The RTD and resistance ranges indicate the voltage conversion
value when measurement current applied.

For integration time is 16.67 ms or more: 120 dB or more

For integration time is 1.67 ms or more: 80 dB or more

(50/60 Hz ±0.1%, 500 Ω unbalanced, between the minus terminal
and ground)

Common-mode voltage between channels (Maximum noise voltage between channels):

250 VACrms (50/60 Hz), reinforced (double) insulation

Noise rejection:

Rejection by the integrating A/D converter and the use of low pass
filters

Input resistance:

DC voltage at 1 V range or less, thermocouple range: 10 MΩ or more

DC voltage at 2 V range or more: Approx. 1 MΩ

Measurement stopped: Approx. 1 MΩ

Insulation resistance: 20 MΩ or more (500 VDC) between the input and earth
Input bias current:

10 nA or less (except for burnout setting)

Withstand voltage:

2300 VAC (50/60 Hz) for one minute between input terminals

3700 VAC (50/60 Hz) for one minute between the input terminal
and earth terminal

Input signal source resistance:

DC voltage and thermocouple: 2 kΩ or less

RTD 50 Ω and 100 Ω systems: 10 Ω or less per cable

RTD 10 Ω and 25 Ω systems: 1 Ω or less per cable

Thermocouple burnout: Superposed electric current system, detection within the

thermocouple range (detection ON/OFF possible)

Up/Down setting is possible, current approximately 100 nA

2 kΩ or less: normal, 10 MΩ or more: line broken

Effect on measurement accuracy: ±15 µV or less (excluding the
effect of signal source resistance)

Parallel capacity during RTD:

0.01 µF or less

4.4 4-CH, High-Speed Universal Input Module (MX110-UNV-H04) Specifications