Measurement interval, integration time, and filter, Measurement synchronization, Common mode voltage – Yokogawa PC-Based MX100 User Manual

Page 27

Advertising
background image

1-18

IM MX100-01E

The following inputs can be used on MXLOGGER or the MX100/DARWIN API (sold
separately). These inputs cannot be used with the MX100 Standard Software.

Input

Measurement Range Type

Rated Measurement Range

DC voltage

60 mV (high resolution)

0.000 to 60.000 mV

1 V

–1.0000 to 1.0000 V

6 V (high resolution)

0.0000 to 6.0000 V

Thermocouple

PLATINEL

0.0 to 1400.0°C

PR40-20

0.0 to 1900.0°C

NiNiMo

0.0 to 1310.0°C

WRe3-25

0.0 to 2400.0°C

W/WRe26

0.0 to 2400.0°C

N (AWG14)

0.0 to 1300.0°C

XK GOST

–200.0 to 600.0°C

RTD

(Measurement current: 1 mA)

Cu10 at 20°C alpha = 0.00392

–200.0 to 300.0°C

Cu10 at 20°C alpha = 0.00393

–200.0 to 300.0°C

Cu25 at 0°C alpha = 0.00425

–200.0 to 300.0°C

Cu53 at 0°C alpha = 0.00426035

–50.0 to 150.0°C

Cu100 at 0°C alpha = 0.00425

–50.0 to 150.0°C

Pt25 (JPt100 × 1/4)

–200.0 to 550.0°C

Cu10 GE (high resolution)

–200.0 to 300.0°C

Cu10 L&N (high resolution)

–200.0 to 300.0°C

Cu10 WEED (high resolution)

–200.0 to 300.0°C

Cu10 BAILEY (high resolution)

–200.0 to 300.0°C

Pt100 GOST

–200.0 to 600.0°C

Cu100 GOST

–200.0 to 200.0°C

Cu50 GOST

–200.0 to 200.0°C

Cu10 GOST

–200.0 to 200.0°C

Measurement Interval, Integration Time, and Filter

The table below shows the available measurement intervals. The module is equipped
with an integrating A/D converter. The selectable integration time varies depending on
the measurement interval as shown in the table below. For details about the relationship
between noise and integration time, see section 2.7, “Measures against Noise on the
MX100” In addition, the type of noise rejection filter switches as shown in the table below.

Measurement

Interval

Integration

Time

Burnout

Detection

Cycle

Filter

Rejected Noise and Notes

100 ms

1.67 ms

1 s

*1

Rectangular

600 Hz and its integer multiples

*2

200 ms

Measurement

interval

500 ms

16.67 ms

60 Hz and its integer multiples

20 ms

50 Hz and its integer multiples

Auto

Automatically detects the power supply

frequency and set 16.67 or 20 ms

1 s

36.67 ms

Trapezoidal

50 Hz or 60 Hz and their integer multiples

2 s

100 ms

Rectangular

10 Hz and its integer multiples

5, 10, 20, 30, 60 s 200 ms

Cos

Fc = 5-Hz low-pass filter

*1 If the measurement interval is 100 ms, burnout detection is carried out on a single channel during

one measurement interval. Therefore, when starting measurement during the burnout condition or

thereafter, burnout can be detected for up to 10 measurements (approximately 1 second).

*2 Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially

with temperature measurement. In such cases, set the measurement interval to 500 ms or higher, or

use the 4-CH, High-Speed Universal Input Module.

Measurement Synchronization

The module is equipped with a single A/D converter, and measurement is made
sequentially. Therefore, measurements of each channel are not synchronized.

Common Mode Voltage

The common mode voltage between channels in a single module is 120 VACrms when
the signal is applied continuously. The common mode voltage between modules and
between a module and earth is 600 VACrms when the signal is applied continuously.

1.4 Functions of the 10-CH, Medium-Speed Universal Input Module