Chapter 4 specification, 1 common specifications, Common specifications -1 – Yokogawa PC-Based MX100 User Manual

Page 89: Index, Standard operating conditions, Transport and storage conditions, Mechanical specifications, Supported standards

Advertising
background image

4-1

IM MX100-01E

Specification

3

2

1

4

5

Index

Chapter 4

Specification

4.1 Common Specifications

Standard operating conditions

Operating temperature range: 0 to 50°C
Operating humidity range:

20 to 80% RH for 0 to 40°C

10 to 50% RH for 40 to 50°C

Rated supply voltage:

AC power supply, 100 to 240 VAC

Supply voltage range used: AC power supply, 90 to 250 VAC
Power supply frequency:

50 Hz ±2%, 60 Hz ±2%

Power consumption:

Up to approximately 70 VA maximum when six modules
are used

Vibration:

10 to 60 Hz 0.2m/s

2

or less

Shock:

Not allowed

Magnetic field:

400 A/m or less (50/60 Hz)

Position:

Upright and horizontal to ground

Structure:

Not explosion-proof

Usage location:

Indoors

Operating altitude:

2000 m or less

Installation category:

II (per CSA22.2 No. 61010-1)

Pollution degree:

2 (per IEC61010-1 CSA22.2 No. 61010-1)

Transport and Storage Conditions

Environmental conditions during transport and storage
Storage ambient temperature: -25 to 60°C
Storage ambient humidity:

5% to 95% RH (no condensation)

Vibration:

10 to 60 Hz 4.9m/s

2

or less

Shock:

392 m/s

2

or less (packaged condition)

Mechanical Specifications

External dimensions:

Approx. 442 Ч 131 Ч 159 mm (when six slots are attached)

Weight:

Approx. 4.1 kg (maximum total weight)

Installation method:

Desktop or floor, panel mount, or attached to a DIN rail

Material:

Steel plate, aluminum die-cast, molded plastic resin

Supported Standards

CSA:

Conforms to CSA22.2 No. 61010-1, overvoltage category II,
pollution degree 2

UL:

Conforms to UL61010B-1 (CSA NRTL/C)

CE:

EMC directives;

EN61326 Class A, EN61000-3-2,

EN61000-3-3, EN55011 Class A, Group 1

Low voltage directive: EN61010-1; measurement category II,

pollution degree 2

C-Tick:

Conforms to AS/NZS CISPR11 Class A Group1

Installation category II: An index for defining transient overvoltages (includes the rated

impulse withstand voltage, and applies to electrical equipment
supplied with power from fixed installations such as distribution
boards).

Pollution degree 2:

Indicates the degree of adherence by a solid, liquid, or vapor that
reduces the withstand voltage or surface resistance (applies only
to normal indoor atmospheres (with non-conductive pollution).

Measurement category II:

Applies to electrical instruments that are powered by wall outlets
and other fixed installations that are wired to distribution boards,
or measurements of such wires.

Advertising