Log file storage, Information saved to the log file, Log file example – Yokogawa PC-Based MX100 User Manual

Page 23: Cf card specifications

Advertising
background image

1-14

IM MX100-01E

Log File Storage

In addition to the measured data that is saved when the communication with the PC is
disconnected, a log file of information related to the CF card operation and information of
power ON/OFF is saved to the CF card in text format (MX100MLG.TXT). The maximum
size of the log file is 40 KB. Up to 1021 events can be stored. When 1021 events are
exceeded, the event is deleted in order from the oldest information. The log is written
(overwritten to the same file name) to the CF card when you press the access stop
switch (see “Ejecting the CF Card” in section 2.8).

Information Saved to the Log File

• Date/Time when the power is turn ON or OFF.
• Date/Time when the CF card is inserted or ejected.
• Save mode when data is saved (backup) and the start/stop log.
• The file creation and deletion log.
• Media-related errors.
• The CF card formatting log.

Log File Example

Yokogawa DAQMASTER MX100 <Media Info>

Date Time Status Message

70/01/01 00:00:00 Power off

03/01/01 00:00:01 Power on

03/03/25 10:12:13 Format ok

03/03/25 11:14:12 Backup start

03/03/25 11:14:12 Mode rotary

03/03/25 11:14:13 Delete (--)

03/03/25 11:14:21 Create 32500000

03/03/25 11:14:36 Code 12 MF

03/03/25 11:14:36 Error P3

03/03/25 11:14:54 Backup stop

03/03/25 11:15:18 Create MX100MLG

03/03/25 11:15:22 Card out

03/03/25 11:15:25 Card in

>>

Time indicating that the settings

have been initialized*

Most-recent information

Media error (records the error code)

Termination mark

Time after resetting the internal clock*

* When the MX100 is initialized, the date/time on the MX100 is reset to 1970/01/01 00:00:00

once, then to the default value on the MX100 (2003/01/01 00:00:00). Then, when the PC

software is started and connection to the MX100 is made, the current date/time on the PC is

transmitted to the MX100. The MX100 is reset to the received date/time.

CF Card Specifications

Item

Specification

Capacity

2 GB maximum

Type

Type I or Type II

Format

Supports quick (logical) format. 1 partition format (hard disk format) is possible

only through a command from the PC when the MX operation mode is idle.

File system

FAT or FAT16

1.2 Main Module Functions

Advertising