1.2 Main Module Functions

The main module is the heart of the MX100.

Ethernet port

Power switch

CF card

slot

Power connector or

screw terminal

7-segment LED

Dipswitch*

DATA ACQUISITION UNIT

ETHERNET

POWER

10BASE - T

100BASE - TX

100 - 240V AC

Do not operate without reading safety precaution in users manual.

70VA MAX

50/60Hz

SW

ON

1 2 3 4 5 6 7 8

Functional ground terminal

* Normally, turn all switches ON.

If the /DS option function is

enabled (see section 5.1),

you can initialize settings

(see section 2.6) and perform

other functions.

1 2 3 4 5 6 7 8

ON

Communications

The main module is equipped with one auto-negotiating 10BASE-T/100BASE-
TX Ethernet port. The LEDs at the upper-left and lower-left of the port indicate the
communication status of the Ethernet interface.

Measurement

The main module acquires measured data sampled at specified intervals on each input
module is acquired. Correcting computation, conversion to physical quantities, and other
processes are performed on the acquired measured data, followed by transmission
of measured data to a PC via the Ethernet interface at 100-ms intervals (shortest).
Even if sampled at intervals less than 100 ms (10 ms or 50 ms), the data is transmitted
collectively at 100-ms intervals. In addition, the main module receives output commands
sent from the PC as necessary and generates signal output instructions to the output
modules.

100 ms

100 ms

100 ms

Creation of output data

Output by an output request from the PC

Data sampling when the

measurement interval is 10 ms

Timing when the creation of the

output data starts

Synchronization of Measurements

• Synchronization between modules
If set to the same measurement interval, measurements made by input modules in the

same acquisition unit are synchronized.

• Synchronization between channels
On the 4-CH, High-Speed Universal Input modules and 10-CH, high speed digital

input modules (-D05 and -D24), measurements between channels are synchronized.

On the 10-CH, Medium-Speed Universal Input modules, 30-CH, Medium-Speed

DCV/TC/DI Input Modules, 6-CH Medium-Speed Four-Wire RTD Resistance
Input modules, and 4-CH, Medium-Speed Strain Input modules (-B12, -B35, and
-NDI), measurements are made sequentially one channel at a time. Therefore,
measurements are not synchronized between channels (we can consider them
synchronized within the measurement interval).