Index saving data to the cf card – Yokogawa PC-Based MX100 User Manual

Page 124

Advertising
background image

5-3

IM MX100-01E

Dual Save Functions (/DS Option)

3

2

1

4

5

Index

Saving data to the CF Card

Transmission of measured data to the PC via communications and saving of measured
data to the CF card in the main module can be carried out simultaneously.
The standard function saves data to the CF card only if communication with the PC
is disconnected during recording by the PC software. With the

Dual Save function,

measured data is saved to the CF card during recording by the PC software.
This function is also available when programming applications with the MX100/DARWIN
API (use in combination with the “Backup”recording command and the “GetData”data
acquisition command).

Note

Under normal functionality, saving of data to the CF card is for backup when communications
are disconnected. The following shows the operation.

Backup When Communication Is Cut

Mode

Timeout

(60 s)

(60 s)

Connects (Monitor display)

Data backup

Command

Data

CF Write

CF File

PC Software

Communication

status

Recording

operation

Disconnects

No data displayed

Save on PC

Save on CF

Recording

starts

Recording

stops

Data acquired

and replenished

retroactively

Replenish

lost data

Comm. cut

Comm. restart

Backup

Backup

ON

Idle

Measure

Start

Stop

Backup

OFF

Measure

Idle

(GetData)

5.2 Explanation of Functions