2-10

IM MX100-01E

Module Type

Max. Input Voltage

Max. Common Mode Voltage

Between Channels Input to Ground

DCV/TC/DI input

module

± 10 VDC: Voltage range of 1 VDC

or less, TC, and DI (contact)

120 VACrms

(50/60 Hz)

600 VACrms

(50/60 Hz)

± 120 VDC: Voltage range of 2

VDC or more, and DI (LEVEL)

Four-wire RTD

resistance input

module

± 10 VDC: Voltage range of 1 VDC

or less, RTD, resistance, and DI

(contact)

Voltage:

120 VACrms

(50/60 Hz)

RTD and

resistance:
50 VACrms

(50/60 Hz)

600 VACrms

(50/60 Hz)

± 120 VDC: Voltage range of 2

VDC or more, and DI (LEVEL)

Strain input

module

± 10 VDC

30 VACrms

(50/60 Hz)

-B12 and -B35:

250 VACrms

(50/60Hz)

-NDI: 30 VACrms

(50/60Hz)

Digital input

module

-D05: ± 10 VDC
-D24: ± 50 VDC

-

250 VACrms

(50/60 Hz)

Module Type

Max. Input Voltage

Max. Common Mode Voltage

Between Channels Output to Ground

Analog/PWM

output module

-

-

250 VACrms

(50/60 Hz)

Digital output

module

250 VAC or 250 VDC

-

250 VACrms

(50/60 Hz)

• The MX100 is a measurement category II (IEC61010-1) and installation
category II (CSA1N.61010-1) instrument.

Consider the points indicated below to prevent noise from entering the measurement

circuit. For a description of the measures against noise on the MX100, see section 2.7.

• Keep the measurement circuit away from the power supply cable (power supply

circuit) and ground circuit.

• It is desirable that the object under measurement is not a noise source. However, if

this is not avoidable, insulate the object under measurement and the measurement
circuit. In addition, ground the object under measurement.

• Shielded wires are effective against noise caused by electrostatic induction. As

necessary, connect the shield to the ground terminal of the MX100 (make sure this
does not lead to grounding at two points).

• Twisting the measurement circuit wires at short intervals is relatively effective against

noise caused by electromagnetic induction.

• Make sure to ground the protective ground terminal through a small grounding

resistance (less than or equal to 100 Ω).

When using the reference junction compensation of the MX100 through thermocouple

input, take measures to stabilize the temperature at the terminal section.

• Always close and secure the terminal cover.
• Do not use thick wires with high heat radiation effects (cross-sectional area of 0.5

mm

2

or smaller recommended).

• Keep the ambient temperature consistent. Large temperature fluctuations occur in

such cases as when a fan nearby is turned ON/OFF.

Connecting the input wires in parallel with other instruments may mutually affect

the measured values.

If you need to make a parallel connection:
• Turn OFF burnout.
• Ground each instrument at a single common point.
• Do not turn ON/OFF the instrument while measurement is in progress. It may cause

adverse affects on the other instrument.

Note that RTDs and resistors cannot be connected in parallel.